maintenance/
troubleshooting
414
Synrad Firestar f-Series operator’s manual
Troubleshooting
RF Drive Switch fault
An RF Drive DC fault occurs on power-up when the tube fails to breakdown or a fault occurs in the RF
Driver’s 96-volt switching circuitry. In this case, the Ready LED ashes 3 blinks. If an RF Drive DC fault
occurs, reset the laser by removing DC power to the laser, wait 30 seconds, and then reapply power. On
Keyswitch-equipped lasers, you must also toggle the Keyswitch from OFF/RESET to ON (or apply a ±5–24
VDC Remote Reset/Start Request pulse with the Keyswitch set to ON) When the Ready LED illumi-
nates, lasing is enabled after a ve-second delay. If the RF Drive DC fault reappears, contact SYNRAD or
a SYNRAD Authorized Distributor.
PWM Drive/RF Oscillation faults
New circuitry on the f-Series RF board has eliminated PWM Drive and RF Oscillation faults. The 4-blink
fault indication no longer exists.
VSWR fault
A VSWR fault indicates an impedance mismatch between the tube and RF Driver and occurs when a
PWM Command signal is rst applied. This fault is annunciated by the Shutter LED ashing continuously
until the fault disappears. If a VSWR fault occurs, the VSWR circuit limits the maximum PWM pulse
length to 12 microseconds (µs). This equates to a 6% maximum duty cycle at a 5 kHz frequency or a 24%
maximum duty cycle at a 20 kHz PWM frequency. When the VSWR fault clears, the laser resumes normal
operation without cycling DC power. If the VSWR fault fails to clear or reappears, contact SYNRAD or a
SYNRAD Authorized Distributor.
The leading causes of VSWR failures are environmental conditions, like cold overnight temperatures
while the laser is o. In cases like this, it may take as long as 60 seconds for tickle to breakdown the tube
and begin normal daily operation.
Other laser faults
Each Symptom listed below describes a particular fault condition. For each Symptom, specic causes and
solutions are described in the Possible Causes section.
Symptom:
A remote interlock condition is indicated by the following status LEDs and I/O states:
Remote Interlock LED – Red Interlock Open output – Closed
Temp LED – Green Over Temperature output Open
Ready LED O Laser Ready output – Open
Shutter LED – Blue Shutter Open output – Closed
Lase LED – O Laser Active output – Open

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Synrad Firestar f series and is the answer not in the manual?

Synrad Firestar f series Specifications

General IconGeneral
BrandSynrad
ModelFirestar f series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals