AR 2000 Operators Manual
103
CAUTION: Although the pressure gauge on the cell may to appear to register zero, the
level of pressure in the cell may still be above ambient.
Disconnect the pressure cell from the manifold at the designated disconnection point at bottom of the
angled pipe.
Raise the head and decouple the magnet assembly and the rotor, and remove the Peltier jacket from the
rheometer using the procedures outlined previously in "Step 4: Loading a Sample."
Remove the cup and rotor for cleaning. See "Maintaining the Cell" for details.