110
NOTES
Delay of ~20mSec is required after a LOAD command, prior to sending any additional command.
UVL and OVP level settings clamp sequence-programmed values.
Loading an empty sequence results in error -286,”Data Load Empty”.
Loading a sequence while any sequence is running, results in error -284,”Program Currently
Running”.
[PROGram]:STEP <DSC>
Function Determines if a trigger executes the whole sequence or a single step.
AUTO - When triggered, it creates waveforms consecutively, until the whole
sequence is over.
ONCE - When triggered, it executes a single step.
[PROGram]:STORe <NR1>
Function Stores a sequence in a specified memory cell 1, 2, 3 or 4.
Stored parameters: mode, voltage/current, dwell/time, step, and counter.
NOTE
A delay of ~100mSec is required after the STORE command, prior sending any additional
command.
[PROGram]:WAVE:CURRent <NRf+>,{<NRf+>}
Function Specifies output current points in a wave.
Current points are separated by commas.
The function accepts up to 100 parameters.
The range is limited by the power supply current programming range.
<NR2>,{<NR2>} 5-digits format
WAVE:CURR 2.5,3.0,3.5; WAVE:CURR MAX,2.5,MIN;
NOTE
A delay of ~100mSec is required after the [PROGram]:WAVE:CURRent command if long
sequences are used, prior sending any additional command.