EasyManua.ls Logo

TDK-Lambda GH10-100 - Page 136

TDK-Lambda GH10-100
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
125
SYSTem:RIN:STATe <Bool>
Function
Enables / Disables the internal resistance function.
Parameters
0|1, OFF|ON
Query
SYSTem:RIN:STATe?
Return
<Bool> 0|1
NOTE
The Internal Resistance function can be set to ON only if the Sequencer, Constant Power Limit,
Analog Programming and Slew-Rate modes are disabled.
SYSTem:REMote[:STATe] <DSC>
Function Sets the power supply control source (front panel or communication) to
local, remote, or LLO mode.
Local Enables the front panel control.
Remote Disables the front panel settings change.
LLO same as remote + disable front panel unlock by the front panel.
Deactivates LLO only by communication or AC recycle.
Parameters
0|1|2 or LOC|REM|LLO
Query
SYSTem:REMote[:STATe]?
Return
<DSC> LOC|REM|LLO
NOTES
System state changes from LOC to REM only if a communication command, which changes a
setting, is sent (Queries does not change system state to Remote mode). System state changes
from LLO to REM after AC power recycle.
SYSTem:SENSe[:STATe] <DSC>
Function
Selects the output sensing point - local or remote
Parameters
0|1, LOC|REM, OFF|ON
Query
SYSTem:SENSe[:STATe]?
Return
<DSC> LOC|REM
SYSTem:SLEW[:STATe] <DSC>
Function Sets OFF / Voltage / Current reference slew control.
OFF Slew control is disabled.
VOLT Voltage reference slew control enabled.
CURR Current reference slew control enabled.
Parameters
0|1|2 or OFF|VOLT|CURR
Query
SYSTem:SLEW[:STATe]?
Return
<DSC> OFF|VOLT|CURR

Table of Contents

Related product manuals