What causes transport lost steps error in tecan sunrise RC Laboratory Equipment?
- SStacie PattersonAug 31, 2025
If the transport system on your Tecan Laboratory Equipment has lost steps, it may be because the optical switch within the transport box is broken, or the plate calibration values have been lost. You can see chapter 4.4.6 Plate calibration Duration Test page 4-16.