USER MANUAL
13
3. Impedance measurement
3.1. Insert Test Probe
For safe and accurate measurement, insert precisely the probe provided with unit.
4 Terminal Pin type probe provided to measure Impedance
Fig 1. Impedance Probe
Fig 2. Impedance Input Terminal
Insert Probe to Ω-H, Ω-L terminal.
Figure 5 Impedance Probe
Figure 6 Impedance input terminal
Be careful any physical injury from the sharp end Pin type probe
When measuring a high-voltage battery of 100V or more, there is a danger of electric shock.
After discharging the measurement probe after measurement, proceed to the next
measurement.
3.2. “O” adjustment
For accurate Ohm measurement, proceed “O” Ohm adjustment when change Probe.
[Parallel Pin Type]
[Coaxial Pin Type]
Parallel Tip Type : The protruding side is the source and the other side is the sensor.
Coaxial Pin Type : The outside is the source and the inside is the sensor.
Source Pin is connected to the upper side and Sensor Pin is connected to the metal part
below.