3.6.3. STRING Record
To measure in String mode, write String Lst of the batteries to be measured.
Refer Record explanation.
[String Mode measurement ]
Press button to display Impedance
Press button to select
.
Press
.
String list is displayed (To make String, refer explanation of Record Function)
Press
,
to select String
Press
to select String.
indicates of slot of measurements of selected String
Press
,
to select slot of New measurement or next slot of
existing measurement
Press
to complete String Select
is ON and ready to measure. Check whether Cell number is same as Battery
to be measured. If not same, press
to move Cursor.
Contact Test Probe to Battery terminal. For correct 4 terminal measurement, press both
inside and outside Probe Pin to contact both side to the Battery Terminal .
activates to hold and save .
Continue to measure on rest cells.
String
String
OK
OK
A HOLD
A REC
◀▶Cursor
A HOLD
A REC

Table of Contents