EasyManua.ls Logo

Tekon 950 - STRING Record

Tekon 950
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER MANUAL
21
3.6.3. STRING Record
To measure in String mode, write String Lst of the batteries to be measured.
Refer Record explanation.
[String Mode measurement ]
Press button to display Impedance ①
Press button to select
. ②
Press
.
③
String list is displayed (To make String, refer explanation of Record Function) ④
Press
,
to select String ⑤
Press
to select String. ⑥
indicates of slot of measurements of selected String ⑦
Press
,
to select slot of New measurement or next slot of ⑧
existing measurement
Press
to complete String Select
⑨
is ON and ready to measure. Check whether Cell number is same as Battery ⑩
to be measured. If not same, press
to move Cursor.
Contact Test Probe to Battery terminal. For correct 4 terminal measurement, press both ⑪
inside and outside Probe Pin to contact both side to the Battery Terminal .
activates to hold and save . ⑫
Continue to measure on rest cells. ⑬
String
String
▲
▼
OK
▲
▼
OK
A HOLD
A REC
◀▶Cursor
A HOLD
A REC

Table of Contents

Related product manuals