Specification-2445A/2455A Service
Table
1-1
(cont)
Characteristics
Performance Requirements
AC POWER SOURCE
Source Voltage
Nominal Ranges
115 V
90
V
to
132 V.
230V
180 V
to
250
V.
Source Frequency
48
Hz
to
440
Hz.a
Fuse Rating 2 A, 250 V, AGC/3AG, Fast blow;
or
1.6 A, 250 V, 5 X
20
mm
Quick-acting.
a
Maximum Power Consumption (fully optioned 120
watts
(180 VA). (a)
instrument)
Primary Circuit Dielectric Voltage Withstand Test
1500 V rms, 60
Hz
for
10
seconds without breakdown. (a)
Primary Grounding
Type test
to
0.1
þÿ©maximum. Routine
test
to
check grounding
continuity between chassis ground and protective earth ground.
(a) Performance
requirement not
checked
In
manual.
1-10