7 Maintaining the product
140
7. Click on [Next >] if the probe should also be included in the
configuration or
8. Select Replace Existing Probe if the probe should replace
another probe.
9. Click on [Next] and follow the instructions of the startup wizard.
Note Further configuration is carried out in the same way
as Add New Component, see Adding new components,
page 134.
Pos: 171 /TD/Produ kt instand halten/te sto Saveris/Kalibri erung und Justage @ 0\mod_118 8476360640_79.doc x @ 2882 @ 25555 @ 1
7.3. Calibration and adjustment
Calibration
The comparison of a measured value with the correct value
under specified conditions. This includes documentation of
the deviation, correction of the measurement uncertainty
and creation of the certificate. Calibration does not include
adjustment of the measuring instrument.
Adjustment
A measuring instrument is used to measure a correct,
known value (confirmed by a superordinate body). The
instrument is then adjusted to this value.