8 Tips and assistance
172
Pos: 190 /TD/Übersc hriften/8. Tipps und Hilfe @ 0\mod_11737898 87985_79.docx @ 406 @ 2 @ 1
8 Tips and assistance
Pos: 191 /TD/Tipps und Hilfe/Fragen und Ant worten/testo Saveris/8. 1 FAQ Saveris Fehlerbehebung SBE @ 12\ mod_1339836791259_ 79.docx @ 131503 @ 2 @ 1
8.1. Questions and answers
Question Possible causes / solution
The converter does not
transfer any data to the
base.
The cable connection to the converter
is faulty.
> Remove the power supply and
check whether the Ethernet cable
is connected correctly.
> Reconnect the power supply.
- The converter checks its
configuration and if it detects an
error, it resets all values to the
factory settings.
Pos: 192 /TD/Tipp s und Hilfe/Fragen und An tworten/testo Saveri s/8.2 FAQ Alarmmeldungen S averis Base @ 12\mod_133968 7134056_79.doc x @ 130826 @ 2 @ 1
8.2. Saveris base alarm messages
Alarm message Possible causes/solution
L_CommUp
L_CommApp
Error during USB or Ethernet
initialisation.
> Disconnect all links to the base.
> Reconnect all links.
> Restart the base.
L_GSM
L_GSMMenue
Error during GSM modem
initialisation.
> Check the GSM module's battery
power.
> Restart the base.
L_RF2010Server
L_RF2010IO
L_RF2010MemPool
L_RF2010StreamRip
L_UDPRF2010
Error during radio module initialisation.
Reboot the base. If the problem
persists, contact the Service team.
> Restart the base.
If the problem persists, please contact
testo Service.
L_UIPrio
L_DispDrvUI,
Error loading the UI/display. Reboot
the base.
> Restart the base.