Operation
Instrument Controls Menu
Thermo Electron Corporation Model 42i Low Source Instruction Manual 3-51
Logic State The Logic State screen is used to change the I/O relay to either normally
open or normally closed.
•Press to toggle and set the logic state between open or closed.
Instrument State The Instrument State submenu allows the user to select the instrument state
that is tied to the selected relay output. A submenu lists signal types of
either alarms or non-alarm to choose from.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State.
OUTPUT RELAY SETTINGS:
>1 NOP GEN ALARM
2 NOP NONE
3 NOP UNITS
4 NOP CONC ALARM
5 NOP NONE
6 NOP NONE
7 NOP NO MODE
ª
RANGE AVG DIAGS ALARM
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
RANGE AVG DIAGS ALARM