4.4 Commands
PM100 / page 41
DATA>
Description: Queries the minimum and the maximum settable wavelength
[meter] for the currently used optical head
Line Frequency Rejection Filter Query
Command syntax:
:FILTER?
Response syntax:
50HZ|60HZ
Description: Queries the current setting for the instrument’s power supply
rejection frequency
Set Line Frequency Rejection Filter Command
Command syntax:
:FILTER 50HZ|60HZ
Response syntax: None
Description: Sets the instrument’s power supply rejection frequency
(50Hz or 60Hz)
Photo Current Query
Command syntax:
:PHOTOCURRENT?
Response syntax:
photo diode current <NR3 NUMERIC RESPONSE DATA>
Description: Queries the last measured value for the photo diode current
[amperes]; works with photodiode sensors only
Photo Voltage Query
Command syntax:
:PHOTOVOLTAGE?
Response syntax:
thermal sensor voltage <NR3 NUMERIC RESPONSE DATA>
Description: Queries the last measured value for the thermal head sensor
voltage [volts]
works with thermal head sensors only
Dark Current Query
Command syntax:
:DARKCURRENT?
Response syntax:
dark current <NR3 NUMERIC RESPONSE DATA>
Description: Queries the set value for the photo diode dark current value
[amperes]; works with Si/Ge sensors only