TH2827 Series Operation Manual Chapter 7 Command Reference
101
Query syntax: CORRection:OPEN:STATe?
Return format: <NR1><NL^END>
The CORRection:SHORt command is used to execute short correction for 48
preset test points (TH2827C has 48 preset test points).
Command syntax: CORRection:SHORt
For example: WrtCmd (“CORR:SHOR”)
The CORRection:SHORt:STATe command is used to set the short correction status.
The CORRection:SHORt:STATe? query returns the current short correction status.
Command syntax
ON
CORRection:SHORt:STATe OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“CORR:SHOR:STAT ON”)
Query syntax: CORRection:SHORt:STATe?
Return format: <NR1><NL^END>
The CORRecition:LOAD:STATe command is used to set load correction. The
CORRection:LOAD:STATe? query returns the current load correction status.
Command syntax:
ON
CORRection:LOAD:STATe OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“CORR:LOAD:STAT ON”)
Query syntax: CORRection:LOAD:STATe?
Return format: <NR1><NL^END>
The CORRection:LOAD:TYPE command is used to set the tested parameter type.
The CORRection:LOAD:TYPE query returns the current parameter type.
Details of function are as follows: