TH2827 Series Operation Manual Chapter 7 Command Reference
67
6.1.5 Delay time
Delay time: time from trigger to start. 0 to 60s are programmable with a
resolution of 1ms.
6.1.6 Connection modes of test terminals
TH2827A adopts 4-terminal test method.
HD(Hcur): current sample high terminal
LD(Lcur): Current sample low terminal
HS(Hpot):voltage sample high terminal
LS(Lpot): Voltage sample low terminal
6.1.7 Test speed (Frequency>=10kHz)
Fast:about 75 times/s(13ms/time)
Medium:about 11 times/s(90ms/time)
Slow:about 2.7 times/s(370ms/time)
The fast and middle speed will be slow down when frequency <10kHz.
6.1.8 Average
Be programmable from 1 to 255.
6.1.9 Display digit
6 digits, max. displayed digit: 999999
6.2 Test signal
6.2.1 Test signal frequency
Test signal is sine wave, accuracy:0.01%
Frequency range:
20Hz~300kHz(TH2827A)
20Hz~500kHz(TH2827B)
20Hz~1MHz (TH2827C)
Min. resolution:0.5mHz