TH2827 Series Operation Manual Chapter 7 Command Reference
75
Table F Direct Calibrated frequency
20 25 30 40 50 60 80 [Hz]
100 120 150 200 250
300
400
500
600
800 [Hz]
1 1.2 1.5 2 2.5
3 4 5 6 8 [kHz]
10 12 15 20 25 30 40 50 60 80 [kHz]
100 120 150 200 250
300
400
500
600
800 [kHz]
1 [MHz]
Cable length factor K
d
0.15%×(F
m
/ 1MHz)
2
× L
2
F
m
: test frequency [MHz]
L: test cable length [m], respectively 0m, 1m, 2m
Table G Temperature factor K
e
Temperature (℃) 0 - 18 18 - 28 28 - 40
Ke 4 1 4
6.3.9 Accuracy of DCR
A (1+R
x
/5M+16m/R
x
)[%] ±0.2m
When the instrument is in Middle and slow speed, A=0.25
In Fast speed, A=0.5
Where, R
x
is the tested resistance.
6.4 Safety requirement
The instrument is the Ⅰ class safety instrument.
6.4.1 Insulation resistance
Under reference working condition,the insulation resistance between power
terminal and instrument jacket should not be smaller than50M.
Under humility condition, the insulation resistance between voltage terminal and
instrument jacket should not be smaller than 2 M.
6.4.2 Insulation intensity
Under reference working condition, the insulation between the power terminal