TH2827 Series Operation Manual Chapter 7 Command Reference
104
reference of the secondary parameter.
For example: WrtCmd (“CORR:SPOT1:LOAD:STAN 100.7, 0.0002”)
Query syntax: CORRection:SPOT<n>:LOAD:STANdard?
Return format: <NR3><NL^END>
The CORRection:USE:DATA? query returns the OPEN/SHORT/LOAD correction
measurement data of 201 correction spots.
Command syntax: CORRection:USE:DATA?<channel number>
Return format:
<open1 A>,<open1 B>,<short1 A>,<short1 B>,<load1 A>,<load1 B>,
<open2 A>,<open2 B>,<short2 A>,<short2 B>,<load2 A>,<load2 B>,
<open3 A>,<open3 B>,<short3 A>,<short3 B>,<load3 A>,<load3 B>,
……
Where,
<open n A> is NR3 data format and the primary open correction data at the
correction spot n.
<open n B> is NR3 data format and the secondary open correction data at
correction spot n.
<short n A> is NR3 data format and the primary short correction data at
correction spot n.
<short n B> is NR3 data format and the secondary short correction data at
correction spot n.
<load n A> is NR3 data format and the primary load correction data at correction
spot n.
<load n B> is NR3 data format and the secondary load correction data at
correction spot n.
The CORRection:CLEar query is used to clear the correction data of all the
correction spots.
Query syntax: CORRection:CLEar
7.1.15 COMParator subsystem commands
The COMParataor subsystem commands are used to set the bin comparator
function including ON/OFF setting, Limit table setting.
Command tree: