TH2827 Series Operation Manual Chapter 7 Command Reference
105
COMParator [:STATe] ON (1)
OFF (0)
:MODE Absolute TOLerance
Percent TOLerance
SEQuence
:TOLerance :NOMinal <value>
:BIN<n> <low limit>,<high limit>
:SEQuence :BIN <BIN1 low limit>,<BIN1 high limit>,
<BIN2 high limit>,<BIN3 high limit>,
,<BINn high limit>
:Secondary LIMit <low limit>,<high limit>
:Auxiliary BIN ON (1)
OFF (0)
:SWAP ON (1)
OFF (0)
:BIN :CLEar
:COUNt [:STATe]
:DATA?
:CLEar
The COMParator[STATe] command is used to set the comparator function as ON
or OFF. The COMParator[STATe]? query returns the current comparator state.
Command syntax:
ON
COMParator[:STATe] OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“COMP ON”)
Query syntax: COMParator[:STATe]?
Return format: <NR1><NL^END>.
The COMParator:MODE command is used to set the comparator mode. The
COMParator:MODE? query returns the current mode.
Command syntax:
ATOLerance
COMParator:MODE PTOLerance
SEQuence
Where,
ATOLerance means absolute tolerance mode.