TH2827 Series Operation Manual Chapter 7 Command Reference
68
6.2.2 Signal mode
Normal: When testing, on measurement display page, voltage across test
terminals may be smaller than preset voltage.
Constant level: The auto adjustment of internal level makes the voltage of DUT
accordant with preset voltage.
6.2.3 Test signal level
Mode Range Accuracy Resolution
Voltage
Normal
constant
level
5mV
RMS
—10V
RMS
10mV
RMS
—5V
RMS
±10%×preset value+2mV
±6%×preset value+2mV
100V
Current
Normal
constant
current
5OA
RMS
—100mA
RMS
100A
RMS
—50mA
RMS
± 10%×preset value+
10A
RMS
± 6%×preset value+
10A
RMS
1A
6.2.4 Output impedance
30, 50 and 100±2% are selectable.
6.2.5 Monitor for test signal level
Mode Range Accuracy
Voltage
5mV
RMS
—2V
RMS
0.01mV
RMS
—5mV
RMS
±3%×reading+0.5mV
±12%×reading +0.1mV
Current
50A
RMS
—20mA
RMS
0.001A
RMS
—50A
RMS
±3%×reading +5A
±12%×reading +1A
6.2.6 Maximum measurement display range
Parameter Measurement display range
LLk 0.00001H 99.9999kH
C
0.00001pF 9.99999F
ZRXDCR 0.00001 99.9999M
YBG 0.00001s 99.9999S
D 0.00001 9.99999
Q 0.00001 99999.9
Deg -179.999°179.999°
Rad -3.14159 3.14159