TH2827 Series Operation Manual Chapter 7 Command Reference
71
6.3.7 Accuracy of Rs
When D
x
value of tested D0.1
The accuracy of R
s
is given by the formula below
R
se
= X
x
×D
e
[]
X
x
= 2πfL
x
=
xfC
2
1
Where, X
x
is the value of test X with the unit [S].
C
x
is the value of test C with the unit [F].
L
x
is the value of test L with the unit [H].
D
e
is the accuracy of D
F is test frequency
6.3.8 Accuracy factor
Figure A Basic measurement accuracy factor A