EasyManua.ls Logo

Tonghui TH2827 Series - Page 109

Tonghui TH2827 Series
131 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TH2827 Series Operation Manual Chapter 7 Command Reference
103
For example: WrtCmd (“CORR:SPOT1:FREQ 2KHZ”) Set the frequency of
correction spot 1 as 2KHZ.
NOTE: <value> should be ranged from 20HZ to 1MHZ (TH2827C), or
return format will report errors.
Query syntax: CORRection:SPOT<n>:FREQuency?
Return format: <NR3><NL^END>
The CORRection:SPOT<n>:OPEN command is execute the open correction for
the current correction spot.
Command syntax: CORRection:SPOT<n>:OPEN
Where,
<n>:
One of the 201 correction spots.
For example: WrtCmd (“CORR:SPOT1:OPEN”) Execute open correction for the
current correction spot 1.
The CORRection:SPOT<n>:SHORt command is used to execute short correction
for the current correction spot.
Command syntax: CORRection:SPOT<n>:SHORt
Where,
<n>:
One of the 201 correction spots.
For example: WrtCmd (CORR:SPOT1:SHOR) Execute short correction for the
current correction spot 1.
The CORRection:SPOT<n>:LOAD:STANdard command is used to set the
standard reference of the current correction spot. The
CORRection:SPOT<n>:LOAD:STANdard query returns the current standard
reference of the current correction spot.
Command syntax: CORRection:SPOT<n>:LOAD:STANdard <REF. A><REF. B>
Where,
<n>:
One of the 201 correction spots.
<REF. A> can be NR1, NR2 or NR3 data format and taken as the standard
reference of the primary parameter.
<REF. B> can be NR1, NR2 or NR3 data format and taken as the standard

Table of Contents

Related product manuals