1. Before Use
16
5 FACTOR A/B
FACTOR A : Select when setting a single correction factor.
The correction is set in the function mode.
FACTOR B : Select when setting the direct correction function, using other instrument BM-
7A, SC-777 or SR-3. The correction is set in the function mode.
Function Mode
「3.1 Function Mode」
Direct Correction Function
「2.1.4 Reducing Error Among a Plurality of
Measurement Instruments and Performing
Measurement」
This switch can be changed during power ON.
6 NC
This is an empty switch and should always be set to the OPEN side.
7 NC
This is an empty switch and should always be set to the OPEN side.
8 PROTECT/WRITE
PROTECT : Prevents writing to internal memory.
This switch should always be set to the PROTECT side during measurement.
WRITE : Allows writing to internal memory.
This switch should be set to the WRITE side when changing memory contents.
This switch can be changed during power ON.
Request
Put the cover on the internal switches/buttons during measurement to prevent a
light from intruding into instrument. Otherwise accurate measurement result
can not be obtained.
Memo
Memo