4. Communication With a PC
69
4.2.21 WG L Command
The chromaticity area to be applied the area correction factor can be store in the instrument by
this command.
WG[m]L[n] : m is the number of the area correction factor group. 1 to 10
n is the area number. 1 to 5
PC Command BM-7A
WG1L3‗####‗####‗####‗####‗#### + (Cr・Lf)
OK + (Cr・Lf)
####‗####‗####‗####‗#### + (Cr・Lf)
END + (Cr・Lf)
The minimum value of the chromaticity x, the minimum value of the chromaticity y, the maximum
value of the chromaticity x, the maximum value of the chromaticity y and the luminance are input
in order in ####‗####‗####‗####‗####.
4.2.22 WHO/VER/SRL Command
The model name, serial number and program version of the instrument are read out by this
command.
PC Command BM-7A
WHO/VER/SRL + (Cr・Lf)
OK + (Cr・Lf)
Model name/serial number/version
+ (Cr・Lf)
END + (Cr・Lf)
4.2.23 UT Command
The unit of luminance of the instrument can be read out by this command.
C:cd/
PC Command BM-7A
UT + (Cr・Lf)
OK + (Cr・Lf)
Unit + (Cr・Lf)
END + (Cr・Lf)