3. Various Settings
46
3.6 Correction Factor
The number of the currently set correction factor is displayed.
Display Method
「3.1.2 Data/Setting Item Display」
The correction factor is a factor for correcting the measurement value by using it to multiply the
measurement value.
The procedure for changing the factor is as described below.
1
Press the [CHANGE] button.
The following screen is displayed.
2
Press the [ROTATION] button, and select the process to be performed (indicated by "*").
DISP&SET : Selected when checking the value of the registered correction factors
(K01 to K10), or when setting the number of the correction factor to be
used.
「3.6.1 Checking/Setting the Correction Factor」
INPUT : Selected when entering or changing the correction factor.
「3.6.2 Inputting the Correction Factor」
CLEAR : Selected when clearing the correction factor.
「3.6.3 Clearing the Correction Factor」
OFF : Selected when not using a correction factor.
BACK : Selected when returning to the previous screen.
3
Press the [ENTER] button.
The screen corresponding to the selected process is displayed.
The correction factor is not valid when dip switch 5 is set to the FACTOR B side.
The factory settings for KX, KY, and KZ are all 1.
FACTOR NUMBER
Memo
FACTOR
DISP&SET OFF
INPUT *BACK
LEAR