3. Various Settings
56
3.9 RS-232C Parameters
The RS-232C interface parameters are displayed.
Display Method
☞
「3.1.2 Data/Setting Item Display」
This setting is a necessary to set when BM-7A connect with a PC through a RS-232C cable.
The procedure for setting parameter is as described below.
1
Press the [CHANGE] button.
The portion of the parameters that can be changed blink.
2
Press the [ROTATION] button and select the setting contents.
Each time the button is pressed, the display changes as shown below.
Baud rate (Communication speed):
38400 → 2400 → 4800 → 9600 →19200→38400 ……
Length (Data length) : 7 → 8 → 7 …
Parity (Parity bit) : ODD → NONE → EVEN → ODD …
Stop bit (Stop bit) : 1 → 2 → 1 …
・ When there are no contents to be changed, press the [ENTER] button, and move the
cursor to the next parameter.
3
After selection is finished, press the [ENTER] button.
Move the cursor to the next parameter.
4
Repeat steps 2 and 3.
・ After the stop bit has been selected, press the [ENTER] button to display the next
setting item (screen).
Memo
Memo
*RS-232C Parameters*
Baud rate=38400
Length=7 Parity=ODD
Stop bit= 1