4. Communication With a PC
72
BM-7fast Mode
Output data are three kinds of data; measurement conditions, color coordinates/correlated color
temperature and tristimulus values.
-M0 Command-
T*R*XnYnZnUCFn_
x=_#.####_y=_#.####_
X=_#.###E±##_Y=_#.###E±##_Z=_#.###E±##CR
-M1 Command -
T*R*XnYnZnUCFn_
u’=_#.####_v’=_#.####_
X=_#.###E±##_Y=_#.###E±##_Z=_#.###E±##CR
-M2 Command -
T*R*XnYnZnUCFn_
Tc=_#.####_duv=_#.###_
X=_#.###E±##_Y=_#.###E±##_Z=_#.###E±##CR
T* : Response speed TF:FAST TS:SLOW
R* : Measurement range RA:AUTO RM:MANUAL
Xn : X
2
measurement range used in actual measurement n = 1 to 5
Yn : Y
measurement range used in actual measurement n = 1 to 5
Zn : Z
measurement range used in actual measurement n = 1 to 5
UC : Measurement unit, cd/㎡
Fn : Measurement angle F1:.° F2:.° F3:° F4:°