4.1.2 V and Vz – View Card and System
Status
The V command displays internal diagnostics. This information is for
inspection only, and cannot be changed from the TTY. If invoked as Vz,
the various counters are cleared, so that subsequent V commands show
what has accumulated since the last Vz. The view listing displays:
Configuration and Internal Status
---------------------------------
and then prints the following lines:
RVP9 Digital IF Signal Processor V13.1(Pol) IRIS–8.13.1
This line is the revision level of the RVP900 software and IRIS version.
Settings were last saved using V12.3
This line displays which version of RVP900 code was the last to write into
the non-volatile RAM. It is printed only if that last version was different
from the version that is currently running. The information is included so
that a "smart upgrade" can often be done, that is, values that did not exist
in the prior release can be filled in with a guess that is better than merely
taking the factory default.
RVP9 started at: 13:07:33 3 JUN 2012
Current time is: 13:14:03 3 JUN 2012
These lines provide information about when the RVP900 was started,
current system time, and implicitly, the uptime.
CPU–Type: Pentium(R) 4 Hyperthreaded
This line displays processor information.
IPP–Library: libippsw7.so v4.0 4.0.19.77
This line displays information about the Intel libraries used for RVP900
processing.
Diagnostics: PASS
NOTE
RVP900 does not actually begin using the current settings until after the
Q command is entered, so that the processor exits the TTY setup mode
and returns to normal operation.