[SMP_ERRR_0014] During X-Ray exposure, mA Feedback is less than 50% off from
the setting value.
Error “5” is displayed on Inverter Board.
[SMP_ERRR_0015] During X-Ray exposure, mA Feedback is more than 150% off
from the setting value.
Error “6” is displayed on Inverter Board.
9.4.2. Motor Error Message sent from MCU to PC
Command Description
[SMP_ERRR_0021] Error on the operation of Rotator motor
: Check Rotator motor and Photo Sensor operation
[SMP_ERRR_0027] Error on the operation of CEPH Sensor motor
: Check CEPH Sensor motor and Photo Sensor operation
[SMP_ERRR_0029] Error on the operation of Sensor tilting motor
: Check Sensor tilting motor and Photo Sensor operation
[SMP_ERRR_0036] Error on operation of 1
st
Collimator(1-axis) left/right movement
: Check 1
st
Collimator Right/Left motor and Photo Sensor
operation
[SMP_ERRR_0037] Error on the operation of Tube Tilting
: Check Tube Tilting motor and Limit Sensor operation
[SMP_ERRR_0041] Error on the operation of Sensor X-axis motor
: Check Sensor X-axis motor and Photo Sensor operation
9.4.3. Exposure Switch Error Message sent from MCU to PC
Command Description
[SMP_ERRR_0060] X-ray Exposure Switch was pressed when the system power
is turned on
9.4.4. Misc. Error Message sent from MCU to PC
Command Description
[SMP_ERRR_0102] CAN communication error(No response)
[SMP_ERRR_0105] Chinrest position error