11.2. Removing the Sensor Sticker
If the measured results exceed the recommended values, measure the radiation dose again after
removing the sensor sticker:
1. Remove the sticker attached in front of the sensor.
2. Attach the dosimeter to the sensor active area.
3. Select a modality (CBCT, PANO, and CEPH) and take an X-ray image.
4. Check the results from the dosimeter.
Sensor sticker
The sticker can be
removed by hands