2.8. CT Number Calibration
To optimize CBCT image quality, the operator should perform the CBCT CT Number Calibration
according to the following procedures.
Make sure that the operator should use TYPE A
during the CT Number Calibration.
(SAP Code - Type A:
M0103175 / Type B: M0104116)
Make sure that the operator should use TYPE A as CT NUMBER CHECK
PHANTOM during the CT Number Calibration.
(SAP Code -Type A: M0103175 / Type B: M0104116)