EasyManua.ls Logo

Vatech A9 PHT-30CSS - Chapter 10. Console Software Error Message List; Hardware Error Code

Vatech A9 PHT-30CSS
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 226 of 244
Chapter 10. Console software error message List
10.1. Hardware Error Code
Error code sending method
Method: Serial communication (RS232)
Format : [SMP_ERRR_####] - #### Error code, tail code only
Classify sub code using Converting function.
Error code output method
[Code : E001.000.000]
Multilingual error phrases output
Main
Code
Sub
Code
Tail
Code
Description Message
E001 ------ --- Hardware related error
000 ---- Success
001 --- [ Generator related error ]
000 - Unknown error
001 - An error that occurs because the
TUBE is not ready
TUBE READY ERROR
002
An error occurred when the cable of
Mono Block and Inverter B'd are not
connected
TUBE MONOBLOCK
ERROR
003 - This error occurs when the current
flows over the maximum allowable
current when driving the inverter (in
case of X-Ray exposure).
TUBE MA OVER ERROR
004
This error occurs when there is a
problem with the tube voltage Ref.
TUBE KV ref ERROR
005 This error occurs when there is a
problem with the tube current Ref.
TUBE mA ref Error
006 -An error occurs when there is a
problem with the tube voltage
feedback.
TUBE KV feedback error
007 -An error occurs when there is a
problem with the tube's current
feedback.
TUBE mA feedback error
008
-An error occurs when the Mono Block
temperature is above the set
temperature.
TUBE TEMP over error
009 -An error occurs when the inverter
output current is abnormal during X-
Ray exposure.
TUBE mA OUTPUT error
010 - This error occurs when the exposure
switch is off after the Exposure on
command.
TUBE EXP ON-
COMMAND ERROR