Page 227 of 244
011 - his error occurs when there is no
exposure off command within 0.5
seconds after the exposure switch is
turned off during exposure.
TUBE EXP OFF
COMMAND ERROR
012
-This error occurs when kV Feedback
differs by -20kV from the set value
during X-ray exposure.
TUBE kV feedback (-)error
013 -This error occurs when kV Feedback
differs by +20kV from the set value
during X-ray exposure.
TUBE kV feedback (+)error
014 - This error occurs when mA feedback
is 50% or less than the set condition
during X-ray exposure.
TUBE mA feedback (-)error
015
- This error occurs when mA feedback
is more than 150% of the setting
condition during X-ray exposure.
TUBE mA feedback
(+)error
002 ---- [ Motor related error ]
000 - Unknown error
020 -This error occurs during the original
movement of the P-axis motor.
Motor P axis error
021 - This error occurs during the original
movement of the rotor axis motor.
Motor R axis error
022 - This error occurs while moving the
origin of the chinrest up/down axis
motor.
Motor jaw up/down the
error
023 - - This error occurs while moving the
origin of the chinrest Left/Right axis
motor.
Motor jaw left/right error
024 - This is an error that occurs while
moving the origin of the CT horizontal
laser up/down axis motor.
Motor CT H up/down the
error
025 - This error occurs during the Canine
beam adjustment motor origin
movement.
Motor canine axis error
026 - This error occurs while moving the
origin of the CEPH secondary
collimator motor.
Motor CEPH 2nd axis
error
027 - This error occurs while moving the
origin of the CEPH sensor motor.
Motor CEPH sensor axis
error
028 - This error occurs while moving the
headband motor home.
Motor headband axis error
029
- This error occurred during sensor
tilting motor origin movement.
Motor Sensor Tilting axis
error
030 - This error occurs while moving the
origin of the left side of the 4-axis
collimator.
Motor 4 Collimator left axis
error
031 - This error occurs while moving the
origin of the right side of the 4-axis
collimator.
Motor 4 Collimator right
axis error