EasyManua.ls Logo

Vatech A9 PHT-30CSS - Page 228

Vatech A9 PHT-30CSS
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 227 of 244
011 - his error occurs when there is no
exposure off command within 0.5
seconds after the exposure switch is
turned off during exposure.
TUBE EXP OFF
COMMAND ERROR
012
-This error occurs when kV Feedback
differs by -20kV from the set value
during X-ray exposure.
TUBE kV feedback (-)error
013 -This error occurs when kV Feedback
differs by +20kV from the set value
during X-ray exposure.
TUBE kV feedback (+)error
014 - This error occurs when mA feedback
is 50% or less than the set condition
during X-ray exposure.
TUBE mA feedback (-)error
015
- This error occurs when mA feedback
is more than 150% of the setting
condition during X-ray exposure.
TUBE mA feedback
(+)error
002 ---- [ Motor related error ]
000 - Unknown error
020 -This error occurs during the original
movement of the P-axis motor.
Motor P axis error
021 - This error occurs during the original
movement of the rotor axis motor.
Motor R axis error
022 - This error occurs while moving the
origin of the chinrest up/down axis
motor.
Motor jaw up/down the
error
023 - - This error occurs while moving the
origin of the chinrest Left/Right axis
motor.
Motor jaw left/right error
024 - This is an error that occurs while
moving the origin of the CT horizontal
laser up/down axis motor.
Motor CT H up/down the
error
025 - This error occurs during the Canine
beam adjustment motor origin
movement.
Motor canine axis error
026 - This error occurs while moving the
origin of the CEPH secondary
collimator motor.
Motor CEPH 2nd axis
error
027 - This error occurs while moving the
origin of the CEPH sensor motor.
Motor CEPH sensor axis
error
028 - This error occurs while moving the
headband motor home.
Motor headband axis error
029
- This error occurred during sensor
tilting motor origin movement.
Motor Sensor Tilting axis
error
030 - This error occurs while moving the
origin of the left side of the 4-axis
collimator.
Motor 4 Collimator left axis
error
031 - This error occurs while moving the
origin of the right side of the 4-axis
collimator.
Motor 4 Collimator right
axis error