95x Series Operating Manual - May 17, 2022
Page 132 of 155
Responds with the <STRING> setting for the START DIO input
Configures the ABORT DIO input to OFF, HI or LO
Responds with the <STRING> setting for the ABORT DIO input
Configures the INTERLOCK DIO input to OFF, HI or LO
Responds with the <STRING> setting for the INTERLOCK DIO input
Configures the SEQUENCE DIO inputs to OFF, HI or LO
Responds with the <STRING> setting for the SEQUENCE DIO inputs
Configures the PASS DIO output to HI or LO
Responds with the <STRING> setting for the PASS DIO output
Configures the FAIL DIO output to HI or LO
Responds with the <STRING> setting for the FAIL DIO output
Configures the TESTING DIO output to HI or LO
Responds with the <STRING> setting for the TESTING DIO output
Configures the DWELL DIO output to HI or LO
Responds with the <STRING> setting for the DWELL DIO output
Configures the HV PRESENT DIO output to HI or LO
Responds with the <STRING> setting for the HV PRESENT DIO output
Configures if the 95x is to fail on arc detection
Responds with the <BOOL> fail on arc detection setting
Configures how a test sequence may be rerun from the front panel
0 – enable fast rerun
1 – enable fast rerun if pass
2 – disable fast rerun
Responds with the <NR1> fast rerun setting
GATEWAY,<NR1>,<NR1>,<NR1>,<NR1>
Configures the Ethernet Gateway IP Address setting
Responds with the Ethernet Gateway IP address setting as four <NR1>
Configures the GPIB address setting
Responds with the <NR1> GPIB address setting
Configures the HV Detect safety feature
Responds with the <BOOL> HV Detect safety feature setting
IP,<NR1>,<NR1>,<NR1>,<NR1>
Configures the Ethernet IP Address setting
Responds with the Ethernet IP address setting as four <NR1>
Configures the maximum allowed discharge current during DCez, DCW
and DCIR type steps to the <NR3> value in amps, values between 1mA
and 200mA are valid
Responds with the <NR3> value of the maximum allowed discharge
current setting
Configures the enable for minimum loading detection during the ramp
period of DCW and DCIR steps -
0 or N – Minimum loading is not detected
1 or Y – Minimum loading is detected as configured in each individual
test step
Responds with the <BOOL> minimum loading detection enable setting
Configures the RS232 baud rate setting
Responds with the <NR1> RS232 baud rate setting
SUBNET,<NR1>,<NR1>,<NR1>,<NR1>
Configures the Ethernet IP Subnet Mask setting
Responds with the Ethernet IP Subnet Mask setting as four <NR1>