EasyManua.ls Logo

Vitrolife EmbryoScope - Page 8

Vitrolife EmbryoScope
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EmbryoScope time-lapse incubator
- 8 -
INSTALLATION AND MAINTENANCE
Installation and repair of the EmbryoScope time-lapse incubator may only be carried out
by a person certified by Vitrolife. The EmbryoScope time-lapse incubator must remain at
the location where it was installed. If an EmbryoScope time-lapse incubator is
disconnected and/or moved without supervision by a person certified by Vitrolife, the
EmbryoScope time-lapse incubator will no longer be approved for clinical use and the
warranty may be voided.
If the EmbryoScope time-lapse incubator or parts of it are modified, appropriate
inspection and testing must be conducted by a person certified by Vitrolife to ensure
continued safe use.
When cleaning and disinfecting the EmbryoScope time-lapse incubator, always use the
prescribed chemical agents as specified in section 8 of the EmbryoScope user manual.
TRANSPORTATION AND RELOCATION OF THE EMBRYOSCOPE TIME-LAPSE
INCUBATOR
While the EmbryoScope time-lapse incubator is still in the shipping boxes, it must only be
moved by using a lift truck or a pallet lifter. Do NOT open the shipping boxes without the
presence of a person certified by Vitrolife.
When the EmbryoScope time-lapse incubator has been unpacked, it must only be moved
by two persons supporting the device underneath the left and right sides and only under
the supervision of a person certified by Vitrolife.
CONNECTION TO EXTERNAL EQUIPMENT
(EN 60601-1 MEDICAL ELECTRICAL EQUIPMENT PART 1)
External equipment intended for connection to signal input, signal output or other
connectors must comply with the relevant IEC standard (i.e. EN 60601-1:2006 Part 1
for medical electrical equipment). In addition, all such combinations systems must
comply with the standard EN 60601-1:2015 Part 2, General requirements for basic
safety and essential performance. Equipment not complying with EN 60601-1:2006
Part 1 must be kept outside the patient environment, i.e. at least 1.5 m from the patient or
the patient support.
Any person who connects external equipment to signal input, signal output or other
connectors has formed a system and is therefore responsible for ensuring that the
system complies with the requirements of EN 60601-1:2006 Part 1. If in doubt, contact
a qualified medical technician or your local representative.

Table of Contents