• The EmbryoScope time-lapse incubator has been tested and found to comply with the
limits for medical devices set out in the IEC 60601-1-2 Edition 4.0 for electromagnetic
compatibility. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation.
Compliance with IEC 60601-1-2 Edition 4.0 ensures compatibility when the EmbryoScope
time-lapse incubator is placed at a minimum distance to nearby instruments. If the
EmbryoScope time-lapse incubator is placed close to other instruments, it is necessary to
observe that the performance of all instruments remains unaffected by the placement.
The EmbryoScope time-lapse incubator generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to other devices, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of the
following measures:
a) Reorient or relocate the receiving device.
b) Increase the separation between the equipment.
c) Connect the equipment to an outlet on a circuit different from that to which the other
device(s) are connected.
Consult the manufacturer, its representative or its dealer for help.
• The use of accessories, transducers and cables other than those specified, with the
exemption of transducers and cables sold by the manufacturer of the system as
replacement parts for internal components, may result in increased emissions or
decreased immunity of the equipment or system.
• The EmbryoScope time-lapse incubator should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the incubator should be
observed to verify normal operation in the configuration in which it will be used.