WM 67851 01/2021
MEDUVENT Standard EN 159
16 Technical data
Display
4.3" TFT color display
Resolution 480 pixels x 272 pixels
Alarm volume
100 %: > 60 dbA
50 %: > 55 dbA
Standards applied
EN 60601-1
EN 60601-1-2
EN 60601-1-6
EN 60601-1-8
EN 60601-1-12
EN 62366-1
EN 1789
EN 13718-1
EN 794-3
ISO 10651-3
ISO 10993-1
RTCA DO-160 G
MIL-STD 810 G
Applied parts acc. to EN 60601-1
• Ventilation mask
• Tracheal tube
Essential performance features
• Delivery of a ventilation volume or triggering of an alarm state
• Limiting of maximum ventilation pressure
Volume-controlled ventilation modes IPPV, Manual, SIMV (optional), S-IPPV (optional)
Pressure-controlled ventilation modes
•CPAP
• CPAP + PS (optional)
• PRVC + PS (optional)
Monitoring
•Volume
• Pressure
• Respiratory rate
Monitoring parameters
•MVi
•pAw
• f(fsp)
Supply gas O
2
Medical oxygen (100 %)
or
concentrator oxygen (93 % ± 3 %)
Operating pressure range 0.3 bar to 6 bar at maximum 15 l/min STPD
Minimum operating pressure 3 hPa. Not adjustable.
Minimum limit pressure, vacuum,
(Plim min)
10 hPa
The device generates no active vacuum.
Maximum limit pressure (Plim max) 60 hPa
Means of limiting pressure Pressure control
Specification Device