160 EN MEDUVENT Standard
WM 67851 01/2021
16 Technical data
Means of safeguarding the minimum
value
Pressure control
Maximum outlet flow 150 l/min (in BTPS)
Mechanical pressure relief/emergency
air valve
Pressure limitation to < 100 hPa
I:E 1: 2 (fixed), in Manual mode 1: 1
Ventilation rate 5 min
-1
to 40 min
-1
±1 min
-1
Respiratory rate measurement
Measuring range: 3 min
-1
to 140 min
-1
Tolerance:
fsp: ±1 min
-1
ftotal: ±1 min
-1
Inspiratory time 0.5 s to 4 s
Tidal volume 50 ml to 2000 ml (±40 ml or ±20 %) (BTPS)
Respiratory minute volume 1.5 l/min to 20 l/min (BTPS)