48 EN MEDUVENT Standard
WM 67851 01/2021
5 Function check
Connect the test lung to the breathing circuit.
Connect the MEDUtrigger.
3. Select Start.
4. If you want to test the oxygen supply: Select Yes.
or
If you do not want to test the oxygen supply: Select No. The
device skips the oxygen test.
5. Set oxygen supply (5 l/min – 15 l/min) (see “6.7 Supplying
oxygen”, page 66).
6. Select Start.