Prepare to clean the equipment
CAUTION Some cleaning agents are not appropriate for all components of
the device. Use only approved cleaning agents, and observe restrictions
noted for some components in the following table. Using unapproved
cleaning agents may cause damage to components.
CAUTION Do not use bleach solutions of any kind when cleaning metal
electrical contacts. They will damage the device.
Select a cleaning agent from the following table.
Section 1. Approved for all Connex Spot Monitor components
Cleaning agent Additional information
Accel INTERVention
Accel TB
CaviWipes
Clinell® Universal Wipes
Oxiver TB
Sani-Cloth
®
Plus
Super Sani-Cloth
®
70 percent isopropyl alcohol solution Applied to a clean cloth
Section 2. Not approved for all Connex Spot Monitor components
Note The following cleaning agents are NOT approved for cleaning Connex Spot
Monitors equipped with the Braun ThermoScan PRO 6000.
Cleaning agent Additional information
Cleancide
Clorox HealthCare Bleach Germicidal Cleaner
Sani-Cloth
®
Bleach
Not approved for cleaning the display
Super HDQ
®
L10
Dilution rate of ½ oz per gallon of water (1:256) applied to a clean
cloth
Tuffie5 Cleaning Wipes
Virex II (256) Dilution rate of ½ oz per gallon of water (1:256) applied to a clean
cloth
10 percent bleach solution (.5% - 1% sodium hypochlorite) applied to a clean cloth
94
Cleaning requirements Connex
®
Spot Monitor