General tab
Specify the language
When you first access Advanced settings, the General tab appears, displaying the
Language tab.
1. Select a language.
2. Do one of the following:
• To continue in Advanced settings, touch another tab.
• To exit Advanced settings and return to the Home tab, touch Exit.
Specify profile settings
Profiles support multiple clinical workflows.
CAUTION Switching from the Office profile drops the Bluetooth connection. Switching to
the Office profile clears both patient information and current readings. Switching to another
profile retains patient information, but current readings are cleared.
1. In Advanced settings, touch the General > Profiles tabs.
2. Specify settings.
Setting Action/Description
Profile types Select the profile types to display: Intervals, Office, and
Spot.
Primary profile Select the primary profile.
Custom profile name If desired, modify the profile names.
3. Touch Apply.
4. Do one of the following:
• To continue in Advanced settings, touch another tab.
• To exit Advanced settings and return to the Home tab, touch Exit.
Specify date and time settings
1. In Advanced settings, touch the General > Date / Time tabs.
2. Specify settings.
Setting Action/Description
Date format Select a date format for display.
Time zone Displays UTC time zone (read only)
Use Network Time Protocol Enable the monitor to query the correct date and time
from a trusted time server. When enabled, manual date
and time entry on the Settings tab is disabled.
3. Touch Test network connection.
Test results appear for the Time sync test, IP address, and DNS query status.
24
Advanced settings Connex
®
Spot Monitor