2. Pull the board straight out to disconnect the mini USB connector.
Reassembly notes
Kit items:
105937, Bluetooth board and a T-10 Torx screw
Note Reassembly is the reverse of disassembly.
Remove the LCD from the main PCA board
1. Carefully separate the LCD flex cable from its connector on the main PCA board by
pulling the cable straight up.
2. Turn over the main PCA board so that the notch on the board faces you.