2. Remove the board.
Reassembly notes
SpO2 board options are Masimo, Nellcor, and Nonin.
Kit items:
105931, adapter board, Masimo board, and 2 Phillips screws (4-40 0.250 pan steel
zinc)
105932, adapter board, Nellcor board, and 2 Phillips screws (4-40 0.250 pan steel
zinc)
105933, Nonin board and 2 Phillips screws (4-40 0.250 pan steel zinc)
Note Reassembly is the reverse of disassembly for each board.
Remove the communications board
The monitor is configured with one of the following communications
boards: wireless radio, Bluetooth wireless technology, or none. Proceed
witih the disassembly steps that apply to your device.
Remove the wireless radio board from the main PCA board
1. Remove the 2 T-10 Torx screws from the wireless radio board attached to the pump
assembly.