Disassembly and repair
These procedures provide instructions for device disassembly and board removal, as
well as component replacement and reassembly.
Each part’s disassembly instructions might include one or both of the following:
• Reassembly notes: This subsection contains information specific to reassembly. At
a minimum, these notes indicate whether or not reassembly is the reverse of
disassembly. The notes also list service kits of replacement parts where applicable.
• When replacing the component: This subsection contains additional instructions
related to installing a new option or replacement part.
Each disassembly step includes drawings that illustrate the components to be removed.
The reassembly notes could be as short as one or two lines when reassembly is the
reverse of disassembly. When reassembly is more complicated, these notes alert you to
any special care required to complete the repair or installation and sometimes introduce
separate reassembly instructions. Line drawings appear in the reassembly notes only
when they differ from the drawings in the disassembly instructions.
WARNING Electrical shock hazard. Disconnect AC power before opening
the device. Disconnect and remove the battery before proceeding with
disassembly. Failure to do this can cause serious personal injury and
damage to the device.
WARNING Risk of fire, explosion and burns. Do not short-circuit, crush,
incinerate, or disassemble the battery pack.
WARNING Safety risk. Do not attempt to service the device when the
device is connected to a patient.
CAUTION Before disassembling the device or installing options,
disconnect the patient from the monitor, power down the device,
disconnect the AC power cord and any attached accessories (for example,
SpO2 sensors, blood pressure hoses and cuffs, and temperature probes)
from the device.
CAUTION If your device is configured with a SureTemp module, remove
the probe well before disassembly.
CAUTION Perform all repair procedures at a static-protected station.
CAUTION When the device case is opened, regard all parts as extremely
fragile. Execute all procedure steps with care and precision.