Remove the pump assembly
1. Remove the pump manifold from the main PCA board transducers.
2. Pinch and remove the pump assembly connector from the main PCA board.
3. Turn over the main PCA board and remove 2 T-10 Torx screws.
Service manual Disassembly and repair 73