Thermometry tests
Calibration key temperature test for SureTemp Plus
Temperature test Specification ± 0.2°F (± 0.1°C) Actual reading Pass Fail
97.3 °F (36.3°C) 97.1 to 97.5°F 36.2 to 36.4°C °F or °C
SureTemp Plus test
Temperature tested Specification ± 0.2 °F (± 0.1 °C) Actual reading Pass Fail
96.8 °F (36.0 °C) 96.6 °F to 97.0 °F 35.9 °C to 36.1 °C °F or °C
101.3 °F (38.5 °C) 101.1 °F to 101.5 °F 38.4 °C to 38.6 °C °F or °C
105.8 °F (41.0 °C) 105.6 °F to 106.0 °F 40.9 °C to 41.1 °C °F or °C
Braun ThermoScan PRO 6000 test
Temperature tested Specification ± 0.4 °F (± 0.2 °C) Actual reading Pass Fail
96.8 °F (36.0 °C) 96.4 °F to 97.2 °F 35.8 °C to 36.2 °C °F or °C
101.3 °F (38.5 °C) 100.9 °F to 101.7 °F 38.3 °C to 38.7 °C °F or °C
105.8 °F (41.0 °C) 105.4 °F to 106.2 °F 40.8 °C to 41.2 °C °F or °C
120
Functional verification and calibration overview Connex
®
Spot Monitor