Message Possible cause Suggested action
Alarm
priority
CSM battery is not
installed. 1C100E
There is no battery in the monitor Check the monitor for a battery and install
one if it is missing. If the problem is still
present, perform diagnostic checks on the
monitor. If the problem persists, replace the
main PCBA.
Very low
Device shutdown is not
available at this time
System shutdown failure Monitor cannot perform an immediate
shutdown. Disconnect AC power and
remove battery.
Information
No valid files found No valid files were found on the
USB flash drive
Re-insert the USB flash drive with valid
files.
Information
Firmware update
successful.
The software successfully updated Information status message; press OK
button to dismiss.
Information
Audio alarms are off. The monitor alarm audio is off Information status message; press OK
button to dismiss.
Information
Advanced settings
unavailable.
The advanced settings are
unavailable because the monitor
is not in an idle state
Confirm that no sensors are attached to the
monitor, no alarms are active, and there is
no unsaved data in the Spot or Intervals
profile.
Information
USB accessory
disconnected.
The USB device was disconnected
from the monitor
Information status message; press OK
button to dismiss
Information
Advanced settings The Advanced settings code was
entered correctly
Information status message; Exit Advanced
settings to dismiss.
Information
Save not successful. The device configuration or logs
were not saved to the USB device
Information status message; press OK
button to dismiss
Information
Save successful. The device configuration or logs
were saved to the USB device
Information status message; press OK
button to dismiss
Information
Software upgrade is
downloading. Do not
shutdown.
The monitor is downloading a
software update
Information status message; press OK
button to dismiss
Information
Factory reset successful. The monitor was reset to factory
settings
Information status message; press OK
button to dismiss
Information
Factory reset failed.
Custom configuration
file not deleted.
The monitor factory reset has
failed.
Information status message; Exit Advanced
settings to dismiss.
Information
Configuration upload
successful.
The device configuration has
successfully uploaded
Information status message; press OK
button to dismiss
Information
Unable to load
configuration; using
factory defaults.
The device configuration was not
successfully uploaded
Information status message; Exit Advanced
settings to dismiss.
Information
Service manual Appendix 197