Temporary patient surface extension by buttery screw
If the bed is equipped accordingly, the patient surface of the bed can be extended and short-
ened again by 10 cm, 20 cm or 32 cm at the foot end without tools. This is why the foot section
of the bed can be extended and locked in three positions. The foot section can be unlocked
with two buttery screws located on the left and right locking mechanism of the foot section.
You need to insert a protector to close the resulting foot-end gap in this case as well.
To extend the patient surface of the sentida low nursing bed:
▪ Loosen and remove the buttery screws on both sides.
The foot section is now released.
▪ Pull the foot section to the correct position and lock it in place by inserting and tightening the
buttery screws. Align the holes to do so.
To shorten the patient surface of the sentida low nursing bed again, proceed in reverse order.
Item Designation Art� no�
1 For type 10 siderails, without padding, without protector 10-0595
2 For type 09 2-part siderails, without padding 10-0595
3 For type 09 3-part siderails, without padding 10-0595
4 For type 10 siderails, without padding, without protector 10-0595
5 For type 09 2-part siderails, without padding 10-0595
6 For type 09 3-part siderails, without padding 10-0595