94
ENGLISH
If the low-set nursing home bed is operated in facilities with medical supervision,
this position may be enabled only by medical personnel or only „under medical
supervision“ for safety reasons! Enabling is done by pressing the corresponding
button for a minimum of 3 seconds. Due to the intended purpose, a Trendelenburg
position is not required!
The leg elevation function is not used for acute care in emergency/shock positio-
ning in low-set nursing home beds. Primarily, this function only serves to effortlessly
set the leg elevation for therapeutic reasons! To avoid the risk of an unintentio-
nal Trendelenburg position, we strongly recommended that the backrest be raised
approximately one hand width at the same time as the legs are raised. It is import-
ant that the resident‘s head is higher than the orthostatic point of the body.
Attention!
Caution!
Automatic prevention (not available for sentida 5-le and 6-le)
First press the button for automatic prevention and then the up or down button.
/
Automatic comfort seating (not available for sentida 5-le and 6-le)
The bed moves to the preset seating position.
Particularly in connection with a bed extension, a risk of getting caught between the
bed and the floor may arise or the head or foot end may touch down on the floor!
Before adjusting the bed, always move the patient surface to an appropriate height!
The same applies in combination with a linen holder!
If the nursing home bed is operated in facilities with medical supervision, this posi-
tion may be enabled only by medical personnel or only „under medical supervision“
for safety reasons! Due to the intended purpose, a Trendelenburg position is not
required!
Attention!
Caution!
/

Table of Contents