EasyManuals Logo

wtw Multi 3430 User Manual

wtw Multi 3430
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Multi 3430 Operation
49
ba75865e08 06/2010
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
Continuing with two-
point calibration
(Buffer: TEC)
5 Wait for the end of the measurement with stability control or
accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears.
6 If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.
30.09.2009 08:00
30.09.2009 08:00
7 Thoroughly rinse the IDS sensor with distilled water.
8 Immerse the IDS pH sensor in buffer solution 2.
9 Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

Table of Contents

Other manuals for wtw Multi 3430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3430 and is the answer not in the manual?

wtw Multi 3430 Specifications

General IconGeneral
Brandwtw
ModelMulti 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals