Operation Multi 3430
50
ba75865e08 06/2010
Continuing with three-
point to five-point
calibration
(Buffer TEC)
10 Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The calibration display for the next buffer appears.
11 Finish the calibration procedure as a two-point calibration with
<M>.
The calibration record is displayed.
12 Thoroughly rinse the IDS pH sensor with distilled water.
13 Immerse the IDS pH sensor in buffer solution 3.
14 Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.