EasyManuals Logo

wtw Multi 3430 User Manual

wtw Multi 3430
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
Multi 3430 Operation
87
ba75865e08 06/2010
Settings You can configure the automatic storage function with the following set-
tings:
Starting the automatic
storage function
To start the automatic storage function, select continue with <S><T>
and confirm with <MENU/ENTER>. The meter switches to the mea-
sured value display.
ID number 1
Interval 30 s
Duration 180 min
Continue
0d03h00min
0 1d17h33min
Automatic data storage
22.09.2009 08:00
Set entire storing dura-
tion
Max. available storing
duration
Graphical display of the
storage usage
Menu item Possible
setting
Description
ID number 1 ... 10000 ID number for the dataset series
Interval 1 s, 5 s, 10 s,
30 s, 1 min,
5min,
10 min,
15 min,
30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration 1min ... x
min
Storage duration.
Specifies after which time the auto-
matic storage should be terminated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.

Table of Contents

Other manuals for wtw Multi 3430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 3430 and is the answer not in the manual?

wtw Multi 3430 Specifications

General IconGeneral
Brandwtw
ModelMulti 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals